KEEP YOUR CALIBRATION CURRENT! The solar reference cell certification service from VLSI Standards allows users to keep their solar reference cells certified and traceable.

Service Description

The Solar Reference Cell Certification Service is a process by which customers submit their own solar reference cells to be measured and certified. The certification is accredited by NIST and is traceable to both the National Renewable Energy Laboratory (NREL), and to the International System of Units (SI). Each solar reference cell is individually calibrated. Spectral response data is also included.

Each cell will receive a Certificate of Calibration and an individual VLSI Standards serial number for historical reference. The Certificate of Calibration reports the certified values and uncertainties of the measured values, and is written in accordance with ISO 17025. It documents the basis of traceability of the reference cell and can be used as part of a quality program complying with ISO requirements.

Product Specifications

SERVICE SPECIFICATIONS TIER 1

  • Certified quantities
    Isc, Voc, Imax, Vmax, Pmax, Fill factor, Area, and Efficiency.

  • Calibration irradiance
    1000 W/m2 (1 sun)

  • Wavelength range
    300 - 1100 nm

  • Spectrum
    AM1.5 Global

  • Calibration temperature
    25ºC

TIER 2 INCLUDES TIER 1 PLUS

Optional temperature dependence measurement of Isc, Voc and efficiency between 20 deg C and 30 deg C is also available.

PLEASE ENSURE THAT YOUR REFERENCE CELL MEETS THESE REQUIREMENTS

  • Must be silicon.

  • Cell dimension limits
    2" x 2"

  • Solar cell current limit
    3 A

  • The reference cell must be enclosed within a metal case, protected by a window.

  • The reference cell must come with electrical connectors for measurements.